Default Setup - Protek 9216A User Manual

Digital lcr meter
Table of Contents

Advertisement

Table 3-1 Test Setup Parameter Settings for Typical Component Values.
Component
Unknown
Resistor
10 H–1 mH
Inductor
10 pF–400 pF
Capacitor
400 pF–1 F

Default Setup

Holding down the backspace key () key while turning on the power to the meter
causes the meter to have the default setup values as listed in the table below.
Note: All user calibration settings and values will be lost.
To prevent the user calibration settings from being destroyed, after turning on
the meter without pressing the backspace key, press RCL  [0]  ENTER. This
key sequence will reset the meter to the default settings listed in the table below
without affecting user-entered calibration values. See the sections that follow for
more details on the settings listed in the table below.
Value
Meas. Mode
Any
AUTO
< 1 k
R+Q
> 1 k
< 10 H
L+Q
1 mH–1H
> 1H
< 10 pF
C+D
C+D
C+D
> 1 F
C+R or C+D
Table 1-1: 9216A default settings
Setting
Parameter
Frequency
Drive Voltage
Bias
Measurement Rate
Averaging
Range Hold
Equivalent Circuit
Display
Trigger Mode
Binning
Equiv. Circuit
Series
Series
Series
Parallel
Series or parallel
Series
Series
Value
AUTO
1KHz
1.0Vrms.
OFF
SLOW
OFF
OFF
SERIES
VALUE
CONT
OFF
19
Frequency
1 kHz
1 kHz
100 or 120 Hz
100 kHz
10 kHz
1 kHz
100 or 120 Hz
10 kHz
10 kHz
1 kHz
100 or 120 Hz