Troubleshooting

Caution
In a centrally networked system, if the previous patient is discharged only
at the monitor and a new patient's data is transferred to the monitor, the
new patient's name and ID will be overwritten by the central station's
broadcast of the previous patient's name. To prevent this situation from
occurring, the previous patient should be discharged from central station
before the transfer.
The following table outlines the Data Transfer behavior under various
events.
Event
Power Off < 1 min
Power Off > 1 min
Op. Mode Change
Config Mode Change
Patient Discharge or End
Case
Results
Transfer will restart automatically
Transfer is stopped. Collection of new data
must be restarted manually.
Transfer is stopped. Data in module remains
intact.
Transfer is stopped. Data in module remains
intact.
Transfer is stopped. Data in monitor is
deleted. Data in module remains intact.
Troubleshooting
Data Transfer 10-19