Selecting Waves for Central Recorders

Selecting Waves for Central Recorders
You can select two waves to be used for delayed alarm recordings at a
Philips patient care system central recorder. The waves you select can
also be displayed at the central station. With a single channel recorder
you can record the primary wave, or with a dual channel recorder you can
record the primary and secondary waves. The primary wave defines
which wave is the default wave at the central station and at other
bedsides in the overview screen.
When ECG is available it must be selected as the primary wave; it cannot
be deselected.
The procedures for configuring delayed recordings are described in the
section "Recording Functions".
Setting up your Monitor 3-39