Cleaning The Roller On The Four Channel (M1117A) Recorder - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Cleaning the
Roller on
the Four
Channel
(M1117A)
Recorder
Equipment
Required
Directions
Step 3
The first sheet will come out folded, as shown.
Note—
If the paper doesn't move through the roller properly, a possible cause is
that the roller needs to be cleaned. (Be sure your hands are clean before
touching the roller).
Lint-free cloths or wipes
Alcohol solvent that is low in water content. We recommend either
reagent grade denaturated ethanol or isopropanol (90% rather than
70%)
1. Turn the power off.
2. Open the door and remove any paper from the recorder.
3. Wipe the visible part of the roller with a saturated (but not dripping)
cloth. Rotate the roller with clean fingertips to expose the entire
surface and wipe the other side.
4. Remove any excess liquid by wiping with a dry cloth.
Caution
The roller must be completely dry before using the recorder or you can
permanently damage the recording mechanism.
Close the door. The paper should feed
automatically.
Loading Paper
Recording Functions 6-55