Changing the Configuration Set
Procedure
The current Configuration Set for the monitor is highlighted and the
universal settings are displayed at the top of the Task Window.
Wait at least one minute before plugging in or removing any modules
from the rack.
3-54
Setting up your Monitor
•
Remove the plug-in modules from the rack before starting the
procedure.
•
Ask your biomedical engineering department or Philips service
engineer to set Parameter Settings Transfer to OFF.
1. Press the softkey
Set you require. The universal settings for the selected
Configuration Set are displayed.
2. Press the hardkey
Change ConfSet
to select the Configuration
to switch to the selected set.
Confirm