When the module is unplugged, it retains data for at least one hour.
After data is transferred to the monitor, the module is automatically
erased. This is to ensure that the module is empty when it is used for the
next transfer.
Symbols to
As detailed in the table below, some versions of this module now use
Indicate Key
symbols, instead of words, to indicate the function of some keys. If the
Functions
monitor's Reference Manual tells you, for example, to select the Monitor
key, you should press the key marked with the corresponding "monitor"
symbol.
The design change also means that you will now find the module's
product number on the rear of the module's housing, not on the front.
Symbol
Name
Monitor
Module
What is
The module is capable of storing and transferring up to 24 hours of trend
Transferred
data and demographic information. When a transfer is initiated, the
following data is transferred:
The following data is not transferred:
Function
initiates transfer from module to
monitor
initiates transfer from monitor to
module
Continuously monitored parameters (for example, ECG)
Resampled Values
Aperiodic measurements (for example, CO)
All patient demographic information except bed number
The time at which calculations were performed in the
Hemodynamic, Oxygenation, and Ventilation calculation task
windows
Marked events
ST events (lead changes, reference changes)
Waveforms, including ST and CO waveforms
Split Screen trends
Arrhythmia events
Data Transfer Module
Which Modules?
Data Transfer Module (M1235A)