Troubleshooting - Emerson Rosemount 2160 Quick Start Manual

Wireless vibrating fork level switch
Hide thumbs Also See for Rosemount 2160:
Table of Contents

Advertisement

Quick Start Guide
Field Communicator
The Network ID and Join Key may be changed in the wireless device by using the
Fast Key sequence in
Table 1. Wireless Setup
Function
Wireless Setup

Troubleshooting

If the device is not joined to the network after power up, verify the correct
configuration of the Network ID and Join Key, and that Active Advertising has
been enabled on the gateway. The Network ID and Join Key in the device must
match the Network ID and Join Key of the gateway.
5.2 Squawk feature
The "Squawk" feature is available to visually locate an individual
Rosemount 2160 amongst multiple networked Rosemount
2160 devices.
Note
The "Squawk" feature requires the Rosemount 2160 to have the optional LCD meter fitted.
AMS instructions
1. From the AMS Device Manager, select the Tag of a Rosemount 2160
to be located.
2. Select Service Tools and then select Maintenance.
3. Select Routine Maintenance.
4. Select Locate Device and then Next to activate Squawk.
5. Look for the pattern 0-0-0-0 on the LCD of the located 'squawking'
Rosemount 2160.
6. To de-activate Squawk, select Next, Cancel, Next, and finally Finish.
Field Communicator instructions
1. From the Home screen, select 3: Service Tools.
2. Select 5: Maintenance and then select 4: Routine Maintenance.
3. Select 1: Locate Device and follow on-screen instructions to activate Squawk.
4. Look for the pattern 0-0-0-0 on the LCD of the located 'squawking'
Rosemount 2160.
5. To de-activate Squawk, exit the feature.
Note
It can take up to 60 seconds for the Rosemount 2160 to return to normal operations.
10
Table
1. Set both Network ID and Join Key.
Key sequence
2, 2, 1
Menu items
Network ID and Join Device to Network
June 2016
0 - 0 - 0 - 0

Advertisement

Table of Contents
loading

Table of Contents