POLARITY 0 / 1

System Settings

This list contains general information on the system, as well as facilities for choosing which protocol to use
for the Modbus Slave output. The System Settings can be accessed by touching the system icon on the
HOME page.
DATE/TIME
IP ADDRESS
SUBNET MASK
GATEWAY
DEVICE ID
BAUD RATE
IS/BURST ENABLE
NEW PASSWORD
REPEAT PASSWORD
FIRMWARE UPDATE

Variables

All variables except average velocity are available to be displayed on the touchscreen display, as live data
with the flowmeter operating in the "Measure" mode. To display the section variables, press the Section
Defines whether the relay is energized or de-energized when it operates as a result
of the event .
Set to 0 for the relay to be normally de-energized.
Set to 1 for the relay to be normally energized.
Note. All relays are de-energized when the electronic unit is without power.
Examples:
To configure a relay to close when the flow in section #1 is negative, set:
TYPE = 1, ASSIGN A SECTION = 1, THRESHOLD = 0, DELAY = 5, POLARITY = 0
To configure a relay to indicate if section #3 fails, or if the power is off, set:
TYPE = 6, ASSIGN A SECTION = 3, THRESHOLD = 0, DELAY = 0, POLARITY = 1
Allows a user to set the current date and time.
Press the value to be changed to bring up the entry pad.
Allow a user to set the flowmeter IP address for TCP/IP communications.
Select Manual to manually set the IP address.
Select Automatic to allow the HMI board to automatically select the available IP
address on a network.
The gray color shows the current selection.
Allow a user to set the subnet mask for TCP/IP communications.
Allow a user to set the gate way for TCP/IP communications.
Allow a user to set the device ID for RS-232 and RS-485 communications.
For TCP/IP communications, this value must be set to 1.
Allow a user to specify the baud rate of the digital output.
9600 is the factory standard.
19200 and 38400 are the other allowable selections.
The gray value will show the selected baud rate.
Allow a user to set the transmit mode.
This would be factory set to IS mode.
Do not change unless instructed to do so by Accusonic.
Allow a user to set a new password for password protection.
A user will be asked to repeat the password for password protection.
Allow a user to update the HMI firmware and/or DSP firmware.
This should not be performed unless specifically instructed to do so by Accusonic.
Specific firmware files are required to perform these updates.
Contact Accusonic for additional information.
User Defined Parameters
7-9