Chapter 2 General Specifications; Transducers; Electronic Unit - Accusonic 8510+ series Reference Manual

Flowmeter
Table of Contents

Advertisement

C H A P T E R 2
General Specifications

Transducers

Temperature Range
Pressure range
Water quality
Characteristic
frequency
Maximum Transmit
Voltage

Electronic Unit

Standard Model
Power Supply
Power Consumption
Temperature Range
Maximum altitude for
normal operation
Dimensions (nom)
Operating: 32F to 105F (0C to 40C)
Storage:
0F to 150F ( –18C to 65C)
Dependent on model
pH: 3.5 to 10
Solids Loading: 0 to 2000 parts per million
Vapors of Ketones & Esters must not be present
1MHz , 500 kHz and 200 kHz using standard electronic filters
Standard Systems: 1100 VDC peak
Flameproof protected transducers: 1100 VDC peak.
Electronic Unit:
90 to 250 Volts AC 47 to 63 Hz
100 to 300 Volts DC without adjustment
Standard Unit: 10 Watts, 20 VA.(with AC power supply)
With relays: 25 Watts, 40 VA. (with AC power supply)
Contact Accusonic for details of models suitable for 24 Volts DC power
supply.
Operating: -4F to 160F (–20C to 70C)
Storage:
0F to 160F (–18C to 70C)
6500 ft. ( 2000 m)
Contact Accusonic for higher altitudes
18 x 16 x 10 inches (460 x 410 x 255 mm)
2-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8510+is series

Table of Contents