Transducer Wiring - Accusonic 8510+ series Reference Manual

Flowmeter
Table of Contents

Advertisement

Unpacking and Installation
4-3
For low voltage d.c. power, observe the correct polarity.
For d.c. power 100 to 300 V d.c. the low potential side of the supply should be connected to the "N"
terminal.

Transducer Wiring

Pull transducer cabling through the appropriate feedthrough and trim each line, leaving enough cable to
reach the transducer terminal blocks at the bottom of the flowmeter console. Tag each cable with a path
number and transducer letter according to the Accusonic numbering convention as shown in Figure 4-2.
Trim the cables, strip back 2 inches of outer sheathing from each, pull inner conductors back from inside
the outer braid, and solder spade lugs to the conductor and shield of each cable as shown in Figure 4-3.
Do not connect the cables to the flowmeter yet. Leave the ends of the cables so that the conductors are not
in contact with one another or with any metal parts on the flowmeter console.