Transducer Cable - Accusonic 8510+ series Reference Manual

Flowmeter
Table of Contents

Advertisement

2-2
Accusonic Model 8510+ Series Flowmeter
Weight
Enclosure protection
Acoustic Paths
Permanent Data
Display
Level Inputs
Analog Outputs
(Isolated)
Alarm relays
(Optional 6)
Integral Data
Logger

Transducer Cable

Unbalanced mode
Balanced mode
Max length between
30 lb (14 kg)
NEMA 4X (IP66)
Up to 10 total, allocated between all the Sections.
Length Standard range
Length Extended ranges
LCD Touchscreen: (800 x 480 pixel)
Character height: 9mm
4 (2-Path & 4-Path Versions) or 8 (10-Path Version)
Useable input range with standard 100 load: 4 -20 mA, (0.4 - 2.0V)
Maximum load (optional): 150 (0.6 - 3V range)
Maximum voltage relative to ground for operation:  20 V DC
Maximum withstand voltage relative to ground: 240 V rms.
Power supply for external transducers: +24 V DC 0.5 A max
4 (2-Path & 4-Path Versions) or 8 (10-Path Version)
Range: 4-20 mA
Maximum load: 600 24 Volts
Resolution: 0.005 mA (16 bit)
Linearity and Stability: 0.04% 180ppm
Isolation: 2500 V rms. common mode to ground
Protection: ± 50 V DC
For Flow or Level exceedance, Faults or Totalizer.
Normally open contacts. 10 A carrying capacity.
Switching capacity: 0.5 A, 110 V DC L/R = 40 ms
Isolation 2000 V AC
104 different variables (all logged)
Interval selectable between 1 second and 24 hours
Capacity of main memory: 8GB minimum
Coaxial RG59 A/U
Special double-jacketed version for underwater use
Twin-axial RG108 for lengths up to 300 ft. (100 m).
Twin-axial RG 22 for long lengths
1600 ft. (500m)
0.7 to 50 ft. (0.2 to 15 m)
5 to 300 ft. (1.5 to 90 m)