Site preparation overview
Introduction
Before the product is delivered, you must prepare the site according to the instructions
in this chapter.
Select the site
Select a site that complies with national and international requirements for biosafety.
General site requirements
The following requirements must be fulfilled at the location where the ImageQuant LAS
500 imager will be installed:
•
•
•
•
•
•
ImageQuant LAS 500 Site Preparation Guide 29-1358-76 AA
WARNING
Spread of biological agents. The operator has to take all necessary
actions to avoid spreading hazardous biological agents. The facility
must comply with the national code of practice for biosafety.
There must not be a heat source such as a radiator near the imager
There must be no risk that the imager can get wet or flooded
The imager must not be exposed to corrosive gas
The imager must not be installed in a dusty environment
The imager must not be installed in a place which is constantly or exessively exposed
to vibration or impacts.
The imager must not be exposed to direct sunlight
2 Site preparation overview
5