Note:
You can take another measurement by pressing the F1 button at
any time although it is recommended that you wait 10 seconds
between readings.
Make sure that the forehead is free of hair and perspiration.
The sensor/probe cover is attached out of the box.
7.

Measuring Temperature via Eardrum

1.
Gently remove the sensor/probe cover to reveal the ear probe.
2.
Insert the ear probe into the ear canal.
3.
Press and release the F2 button. The temperature will be
displayed on the screen instantly.
Note:
Do not press the F2 button for more than 3 seconds as you will
activate the memory retrieval mode.
You can take another measurement by simply pressing the F2
button again.
95.8°F
Incorrect position
Figure 1
For children older than 1 year old, gently pull the ear up to ensure that the
ear canal is straight (See Figure 1). Position the ear probe carefully so
that it is aligned with the eardrum to ensure accurate measurement
readings. This applies to adults as well.
For children younger than 1 year old, gently pull the ear straight back to
ensure that the ear canal is straight (see Figure 2).
98.4°F
Correct Position
7
Figure 2
In order to avoid inaccuracy:
1. The sensor is extremely sensitive. Any debris and/or grease on the
sensor lens might prevent an accurate measurement. For best result,
please check and clean the lens regularly. Please note that the
sensor lens can look deceptively "clean" when it is not. The lens
of a clean sensor should be reflective and gleaming. If it looks
dull and lackluster, then it is dirty. Please clean the lens with a
cotton swab moistened with alcohol and wait 5-10 mins for the
alcohol to completely evaporate before using the thermometer.
2. Please use the thermometer indoor or in an environment where there
is no strong air draft.
3. Do not move during measurement.
4. Make sure there isn't any sweat, water or condensation on the
forehead.
5. If the thermometer is transferred from an environment where the
ambient temperature might be different, please wait at least 10
minutes and follow rule number 2 before using the thermometer.
6. Do not hold the thermometer for a long time as it is highly sensitive to
heat.
7. Keep the infrared sensor probe clean to ensure accurate readings.
8. Before measuring temperature from the ear, remove earwax, if any.
Keep the ear canal clean.