Keithley 2600a series Reference Manual page 578

Sourcemeter
Table of Contents

Advertisement

Section 19: Remote Commands
Table 19-14 (cont.)
KIPulse TSP test script: ConfigPulseIMeasureVSweepLin
TSP project name:
TSP test script name:
Function:
ConfigPulseIMeasureVSweepLin
Usage:
f, msg = ConfigPulseIMeasureVSweepLin
(smu,bias,start,stop,limit,ton,toff,points,buffer,tag[,sync_in]
[,sync_out][,sync_in_timeout][,sync_in_abort])
Parameters
Waveform
19-256
KIPulse
KIPulse
smu,bias,start,stop,limit,ton,toff,points,buffer,tag
[,sync_in][,sync_out])
SourceMeter instrument channel (for example, smua).
smu:
Pulse bias level in amps.
bias:
Pulse sweep start level in amps.
start:
Pulse sweep stop level in amps.
stop:
Voltage limit (for example, compliance) in volts.
limit:
Pulse width (for example, on-time) in seconds.
ton:
Pulse off time in seconds.
toff:
Number of pulse measure cycles.
points:
Reading buffer where pulsed measurements will be stored.
buffer:
If this is nil when the function is called, no measurements
will be made when the pulse train is initiated.
Numeric identifier to be assigned to the defined pulse train.
tag:
Defines a digital I/O trigger input line. If programmed, the
sync_in:
pulse train will wait for a trigger input before executing each
pulse. This parameter is optional.
Defines a digital I/O trigger output line. If programmed, the
sync_out:
pulse train will generate a trigger output immediately prior
to the start of ton. This parameter is optional.
sync_in_timeout: Specifies the length of time (in seconds) to wait for input
trigger (defaults to 10s). New to Firmware 1.3.0
(true or false) Whether to abort pulse if input trigger is not
sync_in_abort:
received before timeout expires (defaults to true). New
to Firmware 1.3.0
step = (stop - start)/(points -1)
step
start
bias
Return to
Section Topics
Series 2600A System SourceMeter® Instruments Reference Manual
step
ton
toff
stop
2600AS-901-01 Rev. B / September 2008

Advertisement

Table of Contents
loading

Table of Contents