Section 19: Remote Commands
smuX.measure.filter.count
Attribute
Number of measured readings to yield one filtered measurement.
Default
1
TSP-Link
This attribute can be accessed from a remote TSP-Link node.
accessibility
Usage
count = smuX.measure.filter.count
smuX.measure.filter.count = count
count
Remarks
• This attribute is the number of measurements that will be performed to yield one filtered
measurement.
• The reset function sets the filter count to 1.
Details
See
Also see
smuX.measure.filter.enable,
Example
Sets filter count for SMU A:
smua.measure.filter.count = 10
smuX.measure.filter.enable
Attribute
Enables/disables filtered measurements.
Default
smuX.FILTER_OFF
TSP-Link
This attribute can be accessed from a remote TSP-Link node.
accessibility
Usage
filter = smuX.measure.filter.enable
smuX.measure.filter.enable = filter
filter
Set filter to one of the following values:
0 or smuX.FILTER_OFF
1 or smuX.FILTER_ON
Remarks
• This attribute enables or disables the filter.
• The reset function disables the filter.
Details
See
Also see
smuX.measure.filter.count,
Example
Enable the filter for SMU A:
smua.measure.filter.enable = smua.FILTER_ON
19-116
Filters
in
Section
6.
Filters
in
Section
6.
Return to
Series 2600A System SourceMeter® Instruments Reference Manual
X = SMU channel (a or b)
-- Reads filter count.
-- Writes filter count.
Set filter count from 1 to 100
smuX.measure.filter.type
X = SMU channel (a or b)
--Reads on/off state of the filter.
--Writes on/off state of the filter.
The filter status.
Disables the filter.
Enables the filter.
smuX.measure.filter.type
Section Topics
.
2600AS-901-01 Rev. B / September 2008