GE pH/C-900 Operating Instructions Manual page 61

Monitor
Table of Contents

Advertisement

pH Measurement
Problem
No response to pH
changes
Small response to pH
changes
Slow pH response or Cali-
bration impossible
Monitor pH/C-900 Operating Instructions 29-0549-25 AB
Corrective action
Check that the electrode cable is connected properly
to the rear of the instrument.
The electrode glass membrane may be cracked.
Replace the electrode.
Clean the pH electrode according to Section 6.4
Cleaning the pH electrode, on page 57 and recali-
brate.
If the problem remains, replace the pH electrode.
Check the pH electrode tip. If it is contaminated
clean the electrode following the instructions in
Section 6.4 Cleaning the pH electrode, on page 57.
If the membrane has dried out, the electrode may
be restored by soaking it in buffer overnight.
Clogged liquid junction. Refer to Section 6.4 Cleaning
the pH electrode, on page 57.
7 Troubleshooting
7.2 Problems and corrective actions
61