DNP v3.00 Device Profile Document

The purpose of this document is to describe the specific implementation of the D istributed
N etwork P rotocol (DNP) 3.0 within the TR-2100 Transient Recorder. This document, in
conjunction with the DNP 3.0 Basic 4 Document Set, and the DNP Subset Definitions
Document, provides complete information on how to communicate with the TR-2100 via the
DNP 3.0 protocol.
The TR-2100 uses the Triangle MicroWorks™, Inc. DNP 3.0 Slave Source Code
Library Version 2.19. This implementation of DNP 3.0 is fully compliant with DNP 3.0
Subset Definition Level 1.
The following sections, in conjunction with the Device Profile Document, provide a
complete interoperability/configuration guide for the TR-2100:
• The Implementation Table
• The Point List Tables
• Configuration methods
DNP V3.00
DEVICE PROFILE DOCUMENT
Vendor Name:
AMETEK Power Instruments
Device Name:
TR-2100 Digital Transient Recorder, using the Triangle MicroWorks, Inc. DNP 3.0 Slave
Source Code Library, Ver 2.19
Highest DNP Level Supported:
For Requests:
For Responses:
Notable objects, functions, and/or qualifiers supported in addition to the Highest DNP Levels Supported (the complete
list is described in the attached table):
For static (non-change-event) object requests, request qualifier code 06 (no range – or all points) is supported.
Maximum Data Link Frame Size (octets):
Transmitted:
Received
mum Data Link Re-tries:
gurable from 0 to 255 via Display Station32
es Data Link Layer Confirmation:
gurable as: Never, Only for multi-frame messages, or Always via DS32
es Application Layer Confirmation:
sending multi-fragment responses
imes
Level 1
Level 1
292
292
Function:
Master
Slave
Maximum Application Fragment Size (octets):
Transmitted: 2048
Received:
2048
Maximum Application Layer Re-tries:
None