Figure 14
3. Open the front panel of the TR-2100 unit using the two top screws, and then turn
the power switch (located in front of the Hard drive) OFF.
4. Next, remove the two white holding bars on either side of the unit. Continue by
removing the gray ribbon cables attached to the Analogue input cards. Now
remove the power cables found on the left hand side of the input cards.
Finally, remove the orange input plugs from the back of each Analogue input
card.
5. The Analogue input cards are now ready to be removed. As noted previously,
the upper two boards are the analogue input boards on a 16-channel unit. If
the unit is a 32-channel recorder, there would be four analogue boards (the
picture above show a 16 channel unit with two analogue input boards and
four digital-only input boards), but the format will remain the same. The
bottom or lowest of the boards is where to start counting. Channel #1 is
located to the right hand side of the top board. Channel #8 can be found on
the top board on the left hand side. Channel #9 can be located on the 2
board from the top, right hand side. Channel #16 can be located on the 2
board on the left hand side. Now you should be able to locate the correct
board to remove from the unit.
6. Once the board has been removed, refer to Figure 15 for jumper locations. In
Figure 15, you will see the Voltage and Current jumpers located across the
input board. The lowest number channel will be located on the right, while
the higher number channel will be on the left.