Hardware Settings - Ametek TR-2100 Operation Manual

Multi - function recorder
Table of Contents

Advertisement

TR-2100 User Manual

Hardware Settings

Acquisition Board Configuration Switches
Set-up the acquisition board configuration switches as shown below. Reference the System
Drawings for number of channels and options to be enabled.
Bank 1
Table 7
Switch
Function
1
Digital-only cards
2
System frequency
3
No. of analogue cards – 1
4
No. of analogue cards – 2
5
Cross triggering
6
GPS receiver
7
Time code port
8
(undefined)
Bank 2
Table 9
Switch
Function
1
Time code port
2
Program port – baud rate
3
4
Printer
5
Modem
6
Synchrophaser on rear
COMM
7
No. of digital only cards -
1
8
No. of digital only cards -
2
Off
On
Switch 3
Yes
No
50 Hz
60 Hz
See 8
Disable
Enable
Yes
No
Master
Slave
* Default setting
Off
On
Switch 2
Disable
Enable
See 10
No
Yes
No
Yes
No
Yes
Switch 7
See 11
Table 8
Switch 4 Analogue Input
Off
Off
Off
On
On
Off
On
On
Table 10
Switch 3
Off
Off
Off
On
On
Off
On
On
Table 11
Switch 8
Digital Input
Off
Off
Off
On
On
Off
On
On
46
Boards
1
2
3
4
Baud Rate
1
2
3
4
Boards
1
2
3
4