Electrical Connection Of The Signal Inputs And Outputs; Connection Diagram Examples - KROHNE Altosonic III Installation And Operating Instructions Manual

3-beam ultrasonic flowmeters for custody transfer of liquid hydrocarbons ufc iii f/...eex ultrasonic flow converter ufs iii f/...eex ultrasonic flow sensor
Table of Contents

Advertisement

The terminal to connect the electrical signal inputs
3.6 Electrical
and outputs consist of 6 connections. For wiring of
connection of
the signal inputs and outputs it is advised to use
the signal
unshielded twisted pairs.
inputs and
outputs
Terminal
P1
P2
S
I/C
The electrical input and output signals can be connected in passive mode. Please observe instrument polarity:
current (I) is always flowing towards P1, P2, S, I/C, terminals (current sink).
3.7 Connection
diagram
examples
Pulse output, P1,
P2
Status output, S
Function
Common ground
Pulse output 1, passive open collector output. Pulse
output to flow computer for volume counting.
Function can be set via menu option.
Pulse output 2, passive open collector output. 90° or
180° phase shifted from P1. For pulse fidelity checking
P1 and P2 should be connected to 2 separate inputs of
a flow computer.
Function can be set via menu option.
Status output. Function can be set via menu option
3.5.0.
Current output (I), 0(4) to 20 mA
Passive open collector current sink output. Digital input
(C)
Function can be set via menu option 3.4.0. and 3.6.0.
For supply:
U ≤ 32V DC, ≤
24V AC
For supply:
U ≤ 32V DC, ≤
24V AC
ALTOSONIC III
Specification
-
I
: 150 mA
max
U
: 32V DC, 24V AC
max
Max frequency: 1,5 kHz
I
: 150 mA
max
U
: 32V DC, 24V AC
max
Max frequency: 1,5 kHz
I
: 150 mA
max
U
: 32V DC, 24V AC
max
Current output (I): I ≤ 22 mA,
R
≤ 680 Ohm. U
= 15V DC.
load
max
Digital input (C): low = 0-5 V DC, high = 15-32 V DC.
Will be switched off when current output activated.
12

Advertisement

Table of Contents
loading

Table of Contents