Installation and
operating instructions
UFM 600 T
ALTOSONIC
Variable area flowmeters
Vortex flowmeters
Flow controllers
Electromagnetic flowmeters
Ultrasonic flowmeters
Mass flowmeters
Level measuring instruments
Communications engineering
Engineering systems & solutions
Switches, counters, displays and recorders
Heat metering
Pressure and temperature

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UFM 600 T ALTOSONIC and is the answer not in the manual?

Questions and answers

Summary of Contents for KROHNE UFM 600 T ALTOSONIC

  • Page 1 © KROHNE 01/94 7.30718.32.00 Installation and operating instructions UFM 600 T ALTOSONIC Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters Level measuring instruments Communications engineering Engineering systems & solutions Switches, counters, displays and recorders Heat metering...
  • Page 2 The contents of this publication are subject to change without notice. This publication is to be used for the standard version only. Thus KROHNE Altometer cannot held responsible for any damage resulting from the incorrect application of this publication to the version actually delivered to you.
  • Page 3: Table Of Contents

    ALTOSONIC Contents UFM 600 T INTRODUCTION ..............6 SYSTEM DESCRIPTION ..........8 Measuring Principle ............8 Measuring System............. 10 Equipment................. 12 Controls and Connections ..........14 Transducers and Mountings ..........16 Display and Data Storage ..........17 Direct Output..............18 Power Supply..............18 OPERATION ..............
  • Page 4 PROGRAMMING MODE..........48 General ................48 How to Program..............49 List of Functions ..............52 Main Menu: 3.0.0 INSTALLATION ........54 Main Menu: 2.0.0 TEST............. 72 Parameter errors ............... 74 MAINTENANCE............... 76 Fuse .................. 76 TECHNICAL DATA............78 General ................78 Transducers &...
  • Page 5 ALTOSONIC List of illustrations UFM 600 T Fig. B.1 : Measuring principle............8 Fig. B.2 : System diagram..............11 Fig. B.3 : UFM 600 T with all equipment displayed ......13 Fig. B.4 : UFC 600 T converter ............15 Fig.
  • Page 7: Introduction

    ALTOSONIC Introduction UFM 600 T System Description The Altosonic UFM 600 T is an ultrasonic flowmeter which can easily be clamped on existing pipelines with inner diameters between 50 mm and 3000 mm (2" to 120") and pipe wall thicknesses up to 40 mm. Measurement is performed obstructionless, without changing the existing pipework and without any headloss.
  • Page 9: Measuring Principle

    ALTOSONIC System Description UFM 600 T Measuring Principle Fig. B.1 : Measuring principle A sound wave that is sent in the direction of the flow through a medium will travel faster than one that is sent in the oppo- site direction. This principle is used in the ultrasonic transit time flowmeter.
  • Page 11: Measuring System

    ALTOSONIC System Description UFM 600 T Measuring System A microprocessor controls the transducers via an analog/ digital interface and calculates the actual flow. The control program is stored in EPROM memory. Parameters for the application and processing the data are entered via the keyboard.
  • Page 12: Fig. B.2 : System Diagram

    SENSOR FLOW SENSOR DOWN SIGNAL EEPROM (µP1) PROCESSOR MICRO RS 232 OUTPUT PROCESSOR (µP2) CURRENT OUTPUT* FREQUENCY* OUTPUT POWER SUPPLY DISPLAY *) galvanically insulated KEYBOARD Fig. B.2 : System diagram...
  • Page 13: Equipment

    ALTOSONIC System Description UFM 600 T Equipment The UFM 600 T flowmeter is fitted in a sturdy case. Figure B.3 shows the UFM 600 T with all items displayed: 1. UFC 600 T signal converter (1x) 2. RS 600 clamp on transducer (2x) 3.
  • Page 14: Fig. B.3 : Ufm 600 T With All Equipment Displayed

    ➉ Fig. B.3 : UFM 600 T with all equipment displayed...
  • Page 15: Controls And Connections

    ALTOSONIC System description UFM 600 T Controls and Connections Figure B.4 shows the controls and connections of the flowmeter unit. 1 Liquid crystal display 2 Keyboard (3 keys) 3 Power supply (term. 11,12) 4 Connector for upstream transducer 5 Connector for downstream transducer 6 RS 232 output for PC 7 mA-Output (term.
  • Page 16: Fig. B.4 : Ufc 600 T Converter

    Fig. B.4 : UFC 600 T converter...
  • Page 17: Transducers And Mountings

    ALTOSONIC System description UFM 600 T Transducers and Mountings (see Fig. B.5) Fig. B.5 : Transducers, signal converter and auxillary items Two mounting rails are clamped against the pipe wall by two straps. The transducers can slide inside the mounting rails to obtain the correct distance between them.
  • Page 18: Display And Data Storage

    Display and Data Storage Display: The following display options are available: - Actual flow rate and direction. - Total positive and negative flow volume since the start of the measuring session. - Absolute flow volume since the start of the measuring session.
  • Page 19: Direct Output

    ALTOSONIC System description UFM 600 T Direct Output Both current and frequency output signals are available for control purposes. These can be either analog signals or just flow direction indicator signals. To avoid adverse effects on equipment connected to the flowmeter a signal hold function can be activated;...
  • Page 21: Operation

    ALTOSONIC Operation UFM 600 T Initial Set-Up The main steps of the installation procedure of the flowmeter are: 1 Install the mounting rails on the pipe. 2 Program the application parameters through the Start Up Menu, up to the point where the microprocessor calculates the transducer distance.
  • Page 22: Installation Of Mounting Rails

    C.1.1 Installation of mounting rails Mounting Considerations: The pipe section on which the transducers will be clamped must always be completely filled with liquid, even if there is no flow. The transducers can be fitted on both horizontal and vertical pipe sections (or any other elevation).
  • Page 23: Fig. C.1 : Mounting Rail

    ALTOSONIC Operation UFM 600 T Mounting Procedure: a Severe rust or thick coatings have to be removed from the pipe where the transducers will be installed. b Slide the securing straps through the slots on both ends of each mounting rail (see fig. C.1). Fig.
  • Page 24: Fig. C.3 : Mounting Positions

    d Use the supplied tape measure to ensure that the rails are fitted diametrically opposite each other (see fig. C.3-1a. ). Distance A1 must equal distance A2. Secure the rails by tightening the straps. Mounting position if calculated transducer distance does not exceed 350 mm (14") Mounting position if calculated transducer distance is between 350 mm (14") and 700 mm (28")
  • Page 25: Programming The Application Parameters

    ALTOSONIC Operation UFM 600 T C.1.2 Programming the application parameters After the flowmeter has been switched on it will start in the measuring mode. Pressing twice the key enters the Start Up Menu. In this menu the application related parameters can be pro- grammed.
  • Page 26 Input of values Input of values outside the range, indicated at the function descriptions, will result in a flashing error message; the mini- mum or maximum value allowed is displayed above the error message. After pressing any key the function is active again and the correct value can be entered.
  • Page 27: Installation Of The Transducers

    ALTOSONIC Operation UFM 600 T C.1.3 Installation of the transducers Since the required transducer distance "S" is known now the transducers can be positioned in the rails: 1 Turn the transducer securing screw completely counter clockwise until the transducer positioning plate touches the top of the transducer housing.
  • Page 28 NOTE: The cable connections at the transducers must always point away from each other (fig. C.5). Make sure that the transducers are not placed over a welding seam. sensor distance S FLOWDIRECTION Fig. C.5 : Upstream and downstream connection 5 Use the cm scale on the mounting rails to position the transducers at the proper distance from each other and hand tighten the securing screws.
  • Page 29: Zero Set / Zero Point Calibration

    ALTOSONIC Operation UFM 600 T C.1.4 Zero set / Zero point calibration Zero Set After leaving the transducer distance function the ZERO SET function can be activated. The flowmeter is functioning now without displaying any flow readings. Only the low signal marker (see C.3.1) gives an indication of the signal quality based on the factory set default zero point calibration.
  • Page 30 Zero Point Calibration Before performing the zero point calibration ensure that: - The unit is functioning properly (see low signal marker). - There is no flow at the location where the transducers are installed and the pipe section is completely filled with liquid. Now activate the Start Up Menu and enter the ZERO SET function: - With...
  • Page 31: Start Up Menu

    ALTOSONIC Operation UFM 600 T Start Up Menu C.2.1 Enter start up menu The Start Up Menu is activated by pressing the key twice from the measuring mode or pressing the key once from the main menu 1.0.0. C.2.2 Application Functions The functions in which the application parameters can be programmed always appear on the display in the same order as described below:...
  • Page 32 PIPE MAT. Program the sonic velocity of the pipe material. The following 5 options are preprogrammed (or will reappear after running 3.5.8 DEFAULT SET, see chapter D): STEEL 3.1700 E 3 m/s ST. STEEL 3.1200 E 3 m/s IRON 2.1200 E 3 m/s 2.1200 E 3 m/s Other 1.5000 E 3 m/s...
  • Page 33 ALTOSONIC Operation UFM 600 T LINING MAT. (does not appear on display when lining = "NO") Program the sonic velocity of the lining material. The following 5 options are preprogrammed (or will reappear after running 3.5.8 DEFAULT SET, see chapter D): 2.1200 E 3 m/s Other 1 1.0000 E 3 m/s...
  • Page 34 according to the user´s requirements (names and sonic velocities). NOTE: If the sonic velocity is not known it can be estimated by using the sonic velocity calculation, see appendix 3. FULL SCALE Select the flow rate units and program the full scale value. A selection can be made out of 9 preprogrammed units and one unit can be specified by the user (see function 3.5.3, 3.5.4 and 3.5.5 in chapter D).
  • Page 35: Plausibility Check

    ALTOSONIC Operation UFM 600 T ZERO SET : Zero point calibration. Select between the factory set default value for the zero point or perform a zero point calibration for the actual situation. See paragraph C.1.4 for the precise conditions and instruc- tions for a calibration.
  • Page 36: Error Detection

    Error Detection C.3.1 Low signal marker function The condition of the input signal from the transducers is indi- cated by a marker in the top left hand corner of the display. The marker consists of four arrows divided over 360 de- grees.
  • Page 37: Error Messages

    ALTOSONIC Operation UFM 600 T C.3.2 Error messages If the microprocessor produces an error message a * sign appears at the lower left hand side of the display. Enter the QUIT / RESET mode to view the error message(s) (see chapter C.5). It depends on the settings of the Installation functions 3.2.4 and 3.2.6, whether the error messages will be displayed in between the flow data.
  • Page 38 STORE: A total number of 350 data blocks are been stored. Rest capacity 40 data blocks till data storage is full. FATAL ERROR: Check sum error detected; contact your supplier. PARAMERROR: This message will not appear during measurement; at the end of both the start up menu and the programming mode a parameter check is per- formed before the measuring mode...
  • Page 39: Data Collection

    ALTOSONIC Operation UFM 600 T Data Collection C.4.1 Data collection procedure Sampling Rate / Data Storage Capacity A total number of 390 data blocks can be stored. This number is independent of the amount of information that is programmed to be on display and/or stored. It is very important to determine a useful combination of the measuring period and the data sampling rate.
  • Page 40 Preparation for new measurements Providing any previous flow data can be discarded, a data collection session can be started as follows: 1 Select the STORE or BOTH option of function 3.6.3 in the programming mode 2 Clear the RAM by means of one of the two ways to Reset Data 3 Start the measuring mode 4 If applicable reset the totalizers...
  • Page 41: Transfer To Pc (Data And Parameters)

    ALTOSONIC Operation UFM 600 T C.4.2 Transfer to PC (data and parameters) Software The floppy disk provided contains programs for the transfer and conversion of the measuring data: - GETFLOW.EXE Function Transfer measuring data and/or parameter settings to the PC The information will be stored in ASCII format.
  • Page 42 Data transfer For the transfer of information from the flowmeter to the PC proceed as follows: Connect the RS 232 cable to the flowmeter and the serial in/output port of the PC (com1 or com2). Enter the programming mode of the flowmeter and activate the function 3.6.1 or 3.6.2 for either the output of the parameter settings or the measuring data;...
  • Page 43: Quit / Reset Menu

    ALTOSONIC Operation UFM 600 T Quit / Reset Menu The measuring mode can be exited with the code as described below to enter the Quit/Reset Menu. In this menu two functions can be used: ERRORLIST: Show the list of errors indicated by the * sign at the left-hand side of the display.
  • Page 44 RESET: In this submenu three reset options are available: TOTAL + Reset + totalizer. TOTAL - Reset - totalizer. RESET DATA Reset all data blocks. In each option YES or NO must be selected. The reset option is to be used to obtain a clear start of a data sampling session.
  • Page 45: Direct Output Ports

    ALTOSONIC Operation UFM 600 T Direct Output Ports (Analog (mA) output and frequency / pulse output) C.6.1 General The options available for these direct output ports are descri- bed further in section D.4 main menu: 3.0.0 INSTALLATION. Refer to the technical specifications if any equipment is to be connected to the current and/or frequency output.
  • Page 46 Active output short-circuit-proof Terminals 4.1/4.2 : for electromechanical (EMC) or electronic (EC) totalizers Terminals 4/4.1/4.2: for electronic (EC) totalizers Amplitude approx. 27 V Load rating see Table "pulse width" Passive output Terminals 4/4.1 open collector for connection of active electronic totalizers (EC) or switchgear Input voltage 5 to 30 V...
  • Page 47: Connection Diagrams

    ALTOSONIC Operation UFM 600 T C.6.5 Connection Diagrams F active (term.5/6) EMC/EC(term.4.1/4.2) 4.1 4.2 4.1 4.2 F actie: F passive EC (term.4/4.1/4.2) EC (term.4/4.1) 4.1 4.2 4.1 4.2 U: 5-30V I max : 100mA...
  • Page 49: Trouble Shooting

    ALTOSONIC Operation UFM 600 T Trouble Shooting The flowmeter does not respond after connection to the main power supply (no display or output messages). Check fuse. The error marker * appears on the display Leave the measuring mode and enter the Quit/Reset menu (press ;...
  • Page 50 replacements. If 3 or 4 markers remain the selection of a diffe- rent location on the pipe wall is re- commended; some irregularities at the interior of the pipe might have caused poor signal reception. Low signal marker: 1 or 2 markers on. Cause Solution Small deviation of the...
  • Page 51: Programming Mode

    ALTOSONIC Programming mode UFM 600 T General In the main menu, two Sub Menus can be activated: Sub Menu 2 "TEST": Four function tests can be executed: Display test Microprocessor test Current output test Frequency output test Sub Menu 3"INSTALLATION": Enter all parameters that determine the initial settings of the flowmeter and the way the data is processed.
  • Page 52: How To Program

    How to Program Push button functions: move the menu pointer one step to the right to activate the submenu displayed. either move the menu pointer to the left to leave a submenu or store entered data and leave the sub- menu.
  • Page 53 ALTOSONIC Programming mode UFM 600 T Menu selection: By entering the main menu, the submenu Start Up 1.0.0. will be on display; use the key to select the submenus TEST 2.0.0. or INSTALL 3.0.0. If the Start Up Menu is to be selected, press key when 1.0.0.
  • Page 54 If any discrepancies are detected the error message PARAMERROR is displayed and a separate menu is activat- ed in which the parameters concerned can be corrected (see section D.6). NOTE: To save the programmed parameters in the Start Up menu (fct.
  • Page 55: List Of Functions

    ALTOSONIC Programming mode UFM 600 T List of Functions 2.00 TEST 2.1.0 DISPLAY 2.1.1 DISP.TEST 2.2.0 PROCESSOR 2.2.1 8048 TEST 2.3.0 CURR. OUTP. 2.3.1 TEST I 2.4.0 FREQ. OUTP. 2.4.1 TEST F 3.0.0 INSTALLATION 3.1.0 FLOW DATA (Submenu) Set flow parameters. 3.1.1 PRIM.
  • Page 56 3.4.0 FREQ OUTP. Set up frequency/pulse output. 3.4.1 FUNCTION F 3.4.2 PULSOUTP. 3.4.3 PULSRATE PULS/UNIT 3.4.4 T- CONST. F 3.5.0 SPEC. FCT. Set special functions. 3.5.1 LANGUAGE 3.5.2 OUTP. HOLD 3.5.3 UNIT TEXT 3.5.4 FACT. QUANT. 3.5.5 FACT. TIME 3.5.6 SET DATE 3.5.7 SET TIME...
  • Page 57 ALTOSONIC Programming mode UFM 600 T Sub Menu: 3.0.0 INSTALLATION set flow parameters 3.1.0 FLOW DATA Primary Correction 3.1.1 PRIM. CORR. This function allows the user to correct the primary constant that is calculated by the microprocessor. Corrections on the primary constant are useful in the case of for example poor installation conditions or comparison to a reference flowme- ter.
  • Page 58: Fig. D.2 : Low Flow Cutoff

    Low Flow Cutoff 3.1.3 L.F. CUTOFF To avoid erroneous measurements at low flow rates the Low Flow Cutoff can be used ( see Fig. D.2). ACTUAL CUTOFF OFF LEVEL FLOW CUTOFF ON LEVEL TIME FLOW METER OUTPUT OUTPUT INACTIVE TIME (flow=0%) Fig.
  • Page 59 ALTOSONIC Programming mode UFM 600 T (only if at 3.1.3 YES was 3.1.4 CUTOFF ON entered) Program the flow rate level below which all outputs will be switched off. Enter the value in % of the full scale. Range: 1- 20 NOTE: This value must be lower than the Cutoff Off value !! (only if at 3.1.3 YES was...
  • Page 60 set output functions 3.2.0 DISPLAY L.C.display NOTE: All functions that are programmed to be on display can also be saved as programmed in function 3.6.3 SEL.COMM. Display Flow rate 3.2.1 DISP. FLOW Determine whether the flow rate should be displayed and if so, in what sort of units;...
  • Page 61 ALTOSONIC Programming mode UFM 600 T unit for totalizers 3.2.3 UNIT TOTAL (only if a totalizer has been set in 3.2.2) Select unit for the totalizers using the key and press to store: Liter US G ------- User defined unit Display Error Messages 3.2.4 ERROR MSG.
  • Page 62 Cyclic Display 3.2.6 CYCL. DISPL. The cyclic display option can be used to automatically change the function of the display, if more than one variable is programmed to be displayed. Only one variable can be shown on the display at a time. With this option active the display sequences from one vari- able to the next every 5 seconds.
  • Page 63 ALTOSONIC Programming mode UFM 600 T set up analog current 3.3.0 CURR. OUTP.I output Program the analog 3.3.1 FUNCTION I current output Select an output option using the key and press to store: 2 DIR Enables flow measurement in 2 direc- tions(+ and -) without change polarity in the output signal (Fig.
  • Page 64: Fig. D.3 : Bi-Directional Flow (I)

    be 0 mA at - 100% flow (fig. D.6) Signal lost Signal received Fig. D.3 : Bi-directional flow (I) Fig. D.4 : Flow direction indication (I) or signal loss indication (I) Fig.D.5 : Positive flow (I) Fig. D.6 : Negative flow and output (I)
  • Page 65 ALTOSONIC Programming mode UFM 600 T Program current output 3.3.2 [I 0 PCT] at 0 %flow Range: 00 mA ≤ [I 0 PCT] ≤ 16 mA NOTE: This value must be lower than the [I 100 PCT] value !! If this is not the case, a PARAMERROR message will appear during the parameter check procedure.
  • Page 66: Fig. D.8: Flow Direction Indication(F)

    3.4.0 FREQ. OUTP. F : set up frequency / pulse output Program the analog fre- 3.4.1 FUNCTION F quency output Select an output option using the key and press to store: 2 DIR Enables flow measurement in 2 direc- tions (fig. D.7). TRAN.
  • Page 67: Fig.d.9: Positive Flow (F)

    ALTOSONIC Programming mode UFM 600 T Fig.D.9: Positive flow (F) 3.4.2 PULSOUTP. : Frequency output mode Select one of the two frequency output modes using the key and press to store: PULSE/UNIT Pulses per unit of volu- PULSRATE Pulses per unit of time at full scale flow.
  • Page 68 or signal lost indication (F) 3.4.3 PULS/UNIT - PULSRATE (pulse/unit or pulse rate depends on selection in 3.4.2) PULSE/UNIT Program number of pul- ses per unit of volume: Pulse per volumetric unit: Range (maximum): PulS/m3 9.9990 E 8 PulS/Liter 9.9990 E 5 PulS/US G 3.7850 E 6 PULSRATE...
  • Page 69 ALTOSONIC Programming mode UFM 600 T set special functions 3.5.0 SPEC. FCT. Display language 3.5.1 LANGUAGE Select a language option using the key and press to store: GB/USA English German French Dutch Hold outputs during 3.5.2 OUTP. HOLD programming If the flowmeter is not in the measuring mode no flow data is available and all output levels will normally drop to zero.
  • Page 70 3.5.4 FACT. QUANT. If a user defined unit is applied the quantity factor has to be programmed. This must be the conversion ratio into m3. In the example with Barrels/Day (see 3.5.3) the quantity unit Barrel has to be converted into m3 (1 Barrel = 1.5898 E-1 m3).
  • Page 71 ALTOSONIC Programming mode UFM 600 T 3.5.8 DEFAULT SET If this function is used all parameters will be reprogrammed to default settings (see Appendix 5). (NOTE: factory settings are not affected by this function) NOTE: The Date, Time and Location data will be lost, too !! This function should only be used to restart programming if the cause of a PARAMERROR can not be found;...
  • Page 72 set RS 232 output / 3.6.0 COMM. DATA store 3.6.1 OUTP. PARAM. With this function all programmed parameters can be sent to a PC via the RS 232 output. NOTE: No error message is given if a communication error (no RS 232 response) occurs! Select YES and press to start the output.
  • Page 73 ALTOSONIC Programming mode UFM 600 T 3.6.4 INTERVAL (not visible if SEL. COMM is OFF) Program the interval between successive data collections as programmed in 3.6.3. Range: 01 - 60 min Location 3.6.5 LOCATION If this function is activated the location, as programmed in the Start Up Menu, is recorded in the header of each data block that is stored or sent to RS 232.
  • Page 75 ALTOSONIC Programming mode UFM 600 T Sub Menu 2.0.0 TEST 2.1.0 DISPLAY 2.1.1 DISPLAY TEST All individual segments of the liquid crystal display are activated in a sequence ending in a flashing display of all segments. No action can be taken. Select YES and press to start the test.
  • Page 76 2.4.0 FREQ. OUTPUT 2.4.1 TEST To check the functioning of both the frequency output and the externally connected instruments and wiring, the frequen- cy output can be set to 5 specific values: 1, 10, 100, 1000 and 10000 Hz The value shown on the display should match the indication on the instrument connected to the frequency output.
  • Page 77: Parameter Errors

    ALTOSONIC Programming mode UFM 600 T Parameter errors If discrepancies in the parameters entered are found the microprocessor will display the message "PARAMERROR" thus preventing you from storing parameters that are incor- rect. The PARAMERROR menu can appear at the end of both the Start Up menu and the Installation Menu.
  • Page 78 4.4.1 L.F. CUTOFF 4.4.2 CUTOFF ON 4.4.3 CUTOFF OFF 4.5.0 F>10KHZ Output frequency is too high, must be less than 10kHz 4.5.1 FULL SCALE 4.5.2 PULSOUTP. 4.5.3 PULSRATE 4.5.4 PULS/UNIT 4.6.0 APPL.DATA Programmed sonic velocities are not within the specified range 4.6.1 PIPE MAT.
  • Page 79: Maintenance

    ALTOSONIC Maintenance UFM 600 T Mains Fuse The mains fuse is located in the connection compartment. NOTE: ONLY USE FUSES AS INDICATED NEXT TO THE FUSE FITTING TO PREVENT SERIOUS DAMAGE TO THE ELECTRONIC EQUIPMENT (see Spare Parts). If the fuse is blown regularly, contact your dealer.
  • Page 81: Technical Data

    ALTOSONIC Technical Data UFM 600 T General Measuring beam 50 ... ≥ 3000 mm (2” ... ≥120”) Diameter range ≤ 40 mm ( ≤ 1.6”) Pipe wall thickness Pipe wall materials metal, plastics, ceramics, asbestos-cement with or without firmly attached linings or coatings Fluids any clean homogeneous liquid with a gaseous or solid particle...
  • Page 82: Transducers & Clamp-On Set

    Transducers & Clamp-on Set Transducers single beam; 2x RS 600 Diameters A single clamp-on set suitable for 50 mm (2”) to 2000 mm (80”) pipes, incorporating 2 rails with cm-scale and straps Materials: Transducer housing nickel plated brass with plastic contact surface.
  • Page 83: Signal Converter

    ALTOSONIC Technical Data UFM 600 T Signal Converter Version UFC 600 T Outputs fully programmable galvanically insulated - analog 0 to 20 mA, or 4 to 20 mA, or I[0%] to I[100%] programmable. ≤ 100 Ohm load - frequency/pulse pulse rate 0.0028 Hz.(= 10 pulses/hour) to 10000 Hz.
  • Page 84 Local display: back-lit LCD Display functions actual flow rate, forward-, reverse- and sum-totalizers (7 digit), transi- tion time and error messages; each programmable for continu- ous or sequential display Display units actual flow liters, m3, or US gallons per sec- ond, minute or hour and one free programmable unit totalizers...
  • Page 85 ALTOSONIC Technical Data UFM 600 T Power supply AC power supply: 85 - 264 VAC DC power supply: 18 - 32 VDC ≤ 10 VA AC Power consumption ≤ 8 W DC Housing Cast aluminium...
  • Page 86 Ambient temp. -25 to +50 °C (-13 to 122°F) Protection category IP 65, acc. to IEC 529 (equivalent to NEMA 4x) Full scale value (Din= inner pipe diameter) lower limit . = 0.5 m/s) 100% min = ( Din [mm]/100 ) * 14.2 100%min = ( Din [inch] )
  • Page 87: Appendix 1 : Pipe Material Sonic Velocities

    ALTOSONIC Appendix UFM 600 T APPENDIX 1 : Pipe Material Sonic Velocities MATERIAL NAME SONIC VELOCITY Carbon Steel c ≤ 0.3% 3064 m/s Carbon Steel c > 0.3% 3173 m/s Carbon-Moly Steel 3173 m/s Cr-Mo-Steel Cr ≤ 3% 3173 m/s Cr-Mo-Steel 5% ≤...
  • Page 88: Appendix 2 : Sonic Velocities Of Liquids

    APPENDIX 2 : Sonic Velocities of Liquids LIQUID NAME SONIC VELOCITY at 20°C Acetone 1174 m/s Benzene 1295 m/s Carbon Disulphide 1149 m/s Caster Oil 1477 m/s Chloroform 987 m/s Diesel Oil 1250 m/s Ethanol 1207 m/s Ethanol Amide 1724 m/s Ethyl Ether 985 m/s Ethylene Glycol...
  • Page 89: Appendix

    ALTOSONIC Appendix UFM 600 T APPENDIX 3 : Sonic velocity calculation Using Indicated Transit Time Install transducers using Sonic Velocity of Water 1500 See upper left corner of the display Look at "Low Signal Marker" to set the correct Transducer Distance Transit Time should be on display, see fct.
  • Page 90: Appendix 4 : Exponential Notation

    APPENDIX 4 : Exponential Notation Most values have to be entered in the exponential notation as will be made clear by the message displayed. Exponential Notation: Y.YYYY E±Z This indicates that Y.YYYY should have a value between 1.0000 and 9.9999 (1.0000 < Y.YYYY < 9.9999). The value of Z gives the multiplication factor, e.g.: Z = +0 indicates Y.YYYY x 1 Z = +1 indicates Y.YYYY x 10...
  • Page 91: Appendix 5 : Default Settings + Example Of Output Parameters

    ALTOSONIC Appendix UFM 600 T APPENDIX 5 : Default Settings + Example of Output Parameters The next page shows all parameters after the function Default Set (3.5.8) has been used. These are the default settings as programmed by the fac- tory.