3. INSTALLATION AND OPERATION
EXCEL
EXCEL
TRANSMITTER
TRANSMITTER
WIRING
WIRING
GREEN/YELLOW (EARTH)
GREEN/YELLOW (EARTH)
Same potential as Excel body and conduit
Same potential as Excel body and conduit
BLACK (0V)
BLACK (0V)
RED (+24V)
RED (+24V)
Shows wires reversed to
disable turbo heating
The earth bonding arrangement must ensure that the maximum peak voltage between the
unit case earth and any field cable conductor is less than 350V. Voltages in excess of this
can cause permanent damage to the unit's RFI protection filters.
Notes:
1.
Only Transmitters have turbo mode heating which can be disabled. The Receiver uses
proportional heating, which cannot be disabled.
2.
Even with turbo mode heating disabled the Transmitter window still receives standard
levels of heating.
3.
Do not reverse the Receiver power supply connections. This will not damage the unit,
but the unit will not function.

3.5 UNPACKING

(1)
Carefully unpack the equipment, observing any instructions that may be printed on or
contained within the packaging.
(2)
Check the contents for damage and against the packing note for deficiencies.
In the event of damage or loss in transit, notify the carrier and Honeywell Analytics or
your local agent immediately.
The gas detector consists of the following items:
Where alternative component sizes are given the size depends on the version of unit,
i.e.mid/long range version or short range version.
1 x Receiver Unit
Honeywell Analytics Junction Box
part number 00780-A-0100
JUNCTION BOX
25
24
CABLE SCREEN
CABLE SCREEN
1 x Transmitter Unit
Searchline Excel
CONTROL
CONTROL
CABINET
CABINET
CONNECTIONS
CONNECTIONS
EARTH
EARTH
+24V
+24V
0V
0V
INSTRUMENT (CLEAN)
INSTRUMENT (CLEAN)
EARTH STAR-POINT
EARTH STAR-POINT