Puma 20/40
DX2-ACT Flash Code List
Note:
These flash codes are the Actuator Module flash codes that are visible on LED 1 of the Actuator Module. They are not the DX
System flash codes that are visible on the DX System status LED on the Master Remote. For DX System flash codes, see DX2
faultfinding table
Sub
Flash
Fault
code
Code
source
(HHP)
00
01
01
User
02
03
none
Slow function
02
battery
00
00
01
02
03
04
05
03
Actuator
06
07
08
All
other
01
02
03
05
06
Configuration
All
other
Meaning
Wizard programming in progress
• Finish programming and then turn the system off and on.
• Make sure that you wait 10 seconds after you have turned the system off before you turn it
on again.
Power Down in progress
• Wait 10 seconds before you turn the system on again.
Joystick source changed to a joystick that is not in the centre position
• Release the joystick to the centre
Invalid Actuator Profile (A-F) or direction demand
• If this happens often, contact the Service Department of Sunrise Medical HCM
see DX2 System condition indications
Battery Voltage too low or too high
• Check the batteries and the cables
• Batteries may be empty: charge the batteries
• Batteries may be damaged: replace the batteries
Batteries may be overcharged: if driving downhill, slow down
Internal current or Voltage fault
• Contact the Service Department of Sunrise Medical HCM
Actuator motor voltage is not what it should be during drive
Possible motor short circuit
• check the actuator cables for damage
• check that the actuator is not faulty
• Otherwise internal controller fault, contact the Service Department of Sunrise Medical HCM
Failed To Stop
Too Many Fast Current Limit Events
Driving Stopped Due To Trip
Internal fault
• Contact the Service Department of Sunrise Medical HCM
Module ID collision
• Check that multiple Actuator Modules in the system do not have the same ID
System check failed
Missing system check
DX System internal fault
Actuators will not move when the DX System itself is in a fault state
Internal DX2-ACT configuration fault
• Check DX BUS cables
• Turn the system off and on several times.
Make sure to wait 10 seconds before you turn the system on again.
• If turning the system off and on does not help, contact the Service Department of Sunrise
Medical HCM
Trouble shooting | 187