Quickie Puma 20 Technical Manual page 184

Powered wheelchair
Hide thumbs Also See for Puma 20:
Table of Contents

Advertisement

184 | Trouble shooting
Trip
Trip Text
Code
User Switch detached
-
Gone to Sleep
Charging
Description
The Omni has detected that the User Switch has become disconnected.
• Check all cables and connectors between the Omni and the User Switch.
If the trip is still present after the above checks have been made, then the User Switch may be
defective. Contact your service agent.
If it is required to use the Omni without a User Switch being connected, then the parameter,
Switch Detect, should be set to Off. If a User Switch is not used the responsibility for that
decision lies with the healthcare professional.
This occurs when the control system has been left inactive for a time greater than the
parameter Sleep Timer.
An entry is made in the system log each time this occurs.
This occurs when the control system detects that a charger is connected to either Inhibit 1
or Inhibit 3. Refer to section 2.3 for connection details The Battery charging screen will be
displayed during charger connection.
An entry is made in the system log each time this occurs.
If an On-Board Charger is used:
• Disconnect the charger from the AC supply.
If an Off-Board Charger is used:
• Disconnect the charger from the Wheelchair.
If the trip is still present after the charger has been disconnected then the
Joystick Module may be defective.