Low P leak (machines with ACGO)

Agent delivery

2067226-001
To run this test, follow the on-screen instructions.
The negative Low P Leak measures machine leaks before the
breathing system, between the common gas outlet and the high-
pressure pneumatics and includes the gas mixer and vaporizer. It
measures low pressure pneumatic leaks with a pass or fail limit of 50
ml.
Note
Any cassette may be used during the check. Using a desflurane
cassette only checks for leaks between the vaporizer and common
gas outlet.
To run this test, follow the on-screen instructions.
The agent delivery test checks the agent delivery systems and the
cassette. This check releases agent to the circuit.
Note
Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
To run this test, follow the on-screen instructions.
5 Preoperative tests