Suction Regulators (Optional) - GE auisys cs2 User's Reference Manual

Digital anesthesia carestation
Table of Contents

Advertisement

Aisys CS²

Suction regulators (optional)

11-22
Venturi Suction Regulator
Performance Category
Supply
Drive Gas Consumption*
Maximum Vacuum*
Maximum Flow*
Vacuum Gauge Accuracy
*Values are approximate.
Continuous Suction Regulator
Performance Category
Supply
Maximum Vacuum*
Maximum Flow*
Vacuum Gauge Accuracy
*Values are approximate.
Pharyngeal Suction
Air or O2 from system gas supply
28 l/min with pipeline drive gas at 280 kPa
52 l/min with pipeline drive gas at 600 kPa
600 mmHg with pipeline drive gas at 280 kPa
550 mmHg with pipeline drive gas at 600 kPa
29 l/min with pipeline drive gas at 280 kPa
32 l/min with pipeline drive gas at 600 kPa
+/- 5% of full scale
Pharyngeal Suction
External vacuum
540 mmHg with external vacuum applied of
540 mmHg and 40 l/min free flow
39 l/min with external vacuum applied of 540
mmHg and 40 l/min free flow
+/- 5% of full scale
2067226-001

Advertisement

Table of Contents
loading

Table of Contents