GE auisys cs2 User's Reference Manual page 254

Digital anesthesia carestation
Table of Contents

Advertisement

Aisys CS²
WARNING
WARNING
13-12
Figure 13-6 • Aladin2 cassette with Easy-Fil system
Do not open or press the cassette filling port or the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
1.
Remove the cassette from the machine, and place it on a
horizontal surface. Make sure that the valve pins on the back of
the cassette do not come in contact with the machine or any
other object.
2.
Align the notches on the bottle adapter to the agent bottle collar,
and tighten the adapter onto the bottle.
To avoid exposure to the anesthetic agent during filling,
make sure that the bottle adapter is properly connected
to the bottle.
--
3.
Remove the filling port cap from the cassette by turning the cap
counterclockwise.
4.
Align the bottle adapter keys with the index slots in the filling
port.
5.
Push the agent bottle firmly into the filling port.
Let the agent flow into the cassette.
Keep the agent bottle aligned with the filling port to prevent
leakage while filling the cassette.
6.
Watch the liquid level indicator.
When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
2067226-001