Malfunction
... the "APrt" and "Time"
menu items are not
visible in the rf menu?
Er:x:11:
... the
appears?
Er:x:12:
... the
appears?
Er:x:16:
... the
appears?
... when the Enter key is
pressed and the
Er:x:71:
display
appears?
... when the Enter key is
pressed and the
Er:x:72:
display
appears?
116 •
• No wireless printer logged on in the wireless
group
- Use the "lrn" menu item to log the wireless
printer into the wireless group (see "Set up
wireless group (Lrn)" on page 109).
The scales are too high, or too great a load is
applied at one point.
display
- Reduce load on scales or distribute weight
more evenly
- Restart scales.
The scales have been switched on with too great
display
a load.
- Reduce load
- Restart scales.
Oscillations have been induced in the scales,
display
preventing determination of the zero point.
- Restart scales
Data transmission not possible, wireless module is
disabled.
- Activate wireless module (see "Activate/
Deactivate wireless module (system)" on
page 112).
Data transmission is not possible, no wireless
group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 109).