4 TEST PROCEDURE: US NAVY PPI 63101-000 (continued)
4.2 TEST PROCEDURE
Measurements are to be made randomly over the prepared surface.
Five measurements should be taken every 90m (1000ft ). Five
measurements should be taken for areas less than 90m (1000ft ).
1
Remove the printed protective
backing and foam centre from the
Bresle patch.
2
Apply the patch to the surface
pressing firmly around the perimeter
of the patch to ensure a complete
seal and remove the clear protective
film cover using the orange tab.
3
Fill a syringe with 3ml of pure water.
4
Insert the syringe into the patch
through the spongy foam perimeter
and inject 1.5ml of pure water into
the patch. Do not remove the
syringe.
5
With the syringe still in the patch,
reposition the needle and evacuate
any air in the patch.
6
Once the air has been removed,
inject the remaining 1.5ml of pure
water.
7
Remove the syringe from the patch.
8
Rub the surface of the patch gently for
10 to 15 seconds to allow the water to
dissolve surface contaminants.
9
Insert the syringe into the patch
through the spongy foam perimeter
and extract the solution from the
patch.
en-5