Transducer; Temperature Sensor; Cabling - Siemens OCM III Instruction Manual

Open channel flow meter
Table of Contents

Advertisement

Battery:

TRANSDUCER

Model:
» XRS-5*
Refer to associated Transducer manual.

TEMPERATURE SENSOR

Model:
» TS-2, LTS-1 or LTS-1C
Refer to associated Temperature Sensor manual.

CABLING

Transducer:
mA Output:
Synchronisation:
Temperature Sensor:
Communication: » RS-232:
» Bipolar Current: » Belden 9552, 2 pair shielded/twisted,
*Note: The XRS-5 must be used with the TS-2 external temperature sensor when
operating with the OCM-3.
Velocity Input:
7ML19985AB01
» 9 V (ANSI / NEDA 1604, PP3 or equivalent) or
3V lithium battery
» RG-62U coaxial
» maximum separation 183 m (600 ft)
» must be run in grounded metal conduit
» Belden 8760 or equivalent
» Belden 8760
» Belden 8760, 1 pair shielded/twisted,
18 AWG or equivalent
» maximum separation 183 m (600 ft)
» can be run in conduit with transducer cable
» Belden 8770, 3 wire shielded, 24 AWG
or equivalent
» maximum separation 15 m (50 ft)
18 AWG or equivalent
» maximum separation 1,500 m (5,000 ft)
» Belden 8760 or equivalent