pH 330i/340i
Single-point calibration
Two-point calibration

4.3.3 ConCal

Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the
buffer solution is to that of the test sample.
For this procedure, use two buffer solutions:
l pH 7.0 ± 0.5
l any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature
sensor.
1
Press the <CAL> key repeatedly until the display ASY and the
function display ConCal appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
2
If necessary, enter the temperature of the first buffer solution
with < > < >.
3
Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4
Press the <RUN/ENTER> key.
The measured pH value appears on the display.
5
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with < > < >.
6
Press the <RUN/ENTER> key. The value of the asymmetry
(mV) and the sensor symbol appear on the display.
7
Press the <RUN/ENTER> key. SLO appears on the display.
Operation
29