pH 330i/340i

4.2.3 Measuring the ORP voltage

In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring
instrument can measure the ORP voltage (U) of a solution.
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the ORP electrode in the test sample.
3
Press the <M> key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the dis-
play.
4
Wait for a stable measured value.
U
pH
ISE
S
1 7
1
0
Time
Baud
2
8
1
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
Note
ORP electrodes are not calibrated. However, you can check ORP elec-
trodes using a test solution.