Measuring The Ph Value - wtw 330i Operating Manual

Handheld ph measuring instrument
Hide thumbs Also See for 330i:
Table of Contents

Advertisement

pH 330i/340i
Setting the resolution
The temperature measurement is absolutely essential for a reproduc-
ible pH measurement. If the measurement is made without a tempera-
ture sensor, proceed as follows:
1
Measure the current temperature of the test sample using a
thermometer.
2
Enter the temperature on the instrument: Set the temperature
value of the test sample using <

4.2.2 Measuring the pH value

1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the <M> key until pH appears on the status display. The
pH value appears on the display.
ISE
U
pH
S
6
9 9
1
Time
Baud
8
2
4
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC ConCal
The measuring instrument shows the pH value with a resolution of 0.01
or 0.001. (Default setting is 0.01). To change the resolution, press the
<M> key while pressing the <RUN/ENTER> key.
> < >.
mol/L
2
mV/pH
% ppm
mg/L
°F
8
%
TP
°C
CalError
TempError
ARng AR
RCL
Operation
15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

340i

Table of Contents