pH 330i/340i
Setting the resolution
The temperature measurement is absolutely essential for a reproduc-
ible pH measurement. If the measurement is made without a tempera-
ture sensor, proceed as follows:
1
Measure the current temperature of the test sample using a
thermometer.
2
Enter the temperature on the instrument: Set the temperature
value of the test sample using <
4.2.2 Measuring the pH value
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the <M> key until pH appears on the status display. The
pH value appears on the display.