Operation
30
Note
At this point, the ConCal calibration can be broken off with <M>. This
corresponds to a single-point calibration. For this, the instrument
uses the Nernst slope (-59.2 mV/pH at 25 °C) and the specified asym-
metry of the electrode.
ISE
U
pH
S
6
S L
1
Time
Baud
8
2
4
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
8
To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9
If necessary, enter the temperature of the second buffer solu-
tion with < > < >.
10
Immerse the pH electrode in the second buffer solution.
11
Press the <RUN/ENTER> key. The second pH value appears
on the display.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with < > < >.
13
When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
mol/L
O
mV/pH
% ppm
mg/L
°F
8
%
TP
°C
CalError
TempError
ARng AR
RCL
pH 330i/340i