Reset - wtw 330i Operating Manual

Handheld ph measuring instrument
Hide thumbs Also See for 330i:
Table of Contents

Advertisement

pH 330i/340i
Measurement
parameters
Configuration
parameters
Resetting the
measuring parameters
4.7

Reset

You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
The following measured parameters (pH InI) are reset to the default
condition:
Measuring mode
Asymmetry
Slope
Calibration procedures
Temperature, manual
Resolution (pH display)
Display during the
pH calibration
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
The following configuration parameters (InI) are reset to the delivery
status:
Baud rate
Interval 1
(automatic save)
Interval 2
(for data transmission)
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <CAL> key.
pH
0 mV
-59.16 mV/pH
AutoCal TEC
25 °C
0.01 (low resolution)
Buffer nominal value
4800
OFF
OFF
Operation
49