Maintenance; Cleaning The Compressor - Pari mini Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

en
- 37 -

4 MAINTENANCE

4.1 Cleaning the compressor

Liquids can conduct, ...2
WARNING
Liquids can conduct electricity, thereby posing a risk of electric
shock. Therefore, always switch the compressor off and
disconnect the power adapter from the mains socket before
starting to clean it.
• Wipe the outer surface of the housing with
a clean, damp cloth.
Liquids that get into the device ...
NOTICE
Liquids that get into the device can cause a fault in the
device. Therefore, never spray any liquids onto the compressor
or the power cord.
If liquid does get into the compressor, the device must not be
If liquid does get into ...
used under any circumstances. Before using the compressor
again, you must contact the PARI GmbH Technical Service
department or the PARI Service Center [See also: 7.4 Contact].
In order to clean the nebuliser, the connection tubing and the
In order to clean ...
accessories, follow the notes in the respective instructions for
use.