Guidance and manufacturer's declaration – electromagnetic
immunity
The BUA5000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BUA5000 should assure that it is used in such an
environment.
IEC 60601 test
Immunity test
level
Electrostatic
±6kV Contact
discharge (ESD)
±8kV Air
IEC 61000-4-2
Radiated RF
3V/m 80MHz to
2.5GHz
IEC 61000-4-3
Conducted RF
3Vrms 150kHz to
80MHz
IEC 61000-4-6
Electrical fast
±2kV
transient
power line
IEC 61000-4-4
±1kV I/O lines
±1kV
Surge
differential
IEC 61000-4-5
±2kV common
Power frequency
magnetic field
3A/m
IEC 61000-4-8
Voltage dips, short
>95% dip 0.5 cycle
interruptions and
voltage variations
60% dip 5 cycles
on power supply
70% dip 25 cycles
input lines
95% dip 5 sec.
IEC 61000-4-11
12
Compliance
Electromagnetic
level
environment – guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
Complies
synthetic material, the relative
humidity should be at least
30%.
Field strengths outside the
Complies
shielded location from fixed
RF transmitters, as determined
by an electromagnetic site
survey, should be less than
3V/m.
Interference may occur in the
vicinity of equipment marked
Not
with the following symbol:
applicable
(no electrical
cabling)
Separation distance
calculation provided above.
If a known transmitter is
present the specific distance
can be calculated using the
equations.
Not
applicable
The BUA5000
is solely battery powered.
Not
applicable
Power frequency
magnetic fields should be at
levels characteristic of a
Complies
typical location in a typical
commercial or hospital
environment.
Not
The BUA5000
applicable
is solely battery powered.