Installing the Refractive Index Detector

Flow Connections to the Detector

Preparations
Parts required
When working with solvents please observe appropriate safety procedures (for
WA R N I N G
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet supplied by the solvent vendor, especially
when toxic or hazardous solvents are used.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
1 Press the release buttons and remove the front cover to
gain access to the interface port area.
42
Detector is installed in the LC system.
Other modules
Interface tubing kit G1362-68706
Interfacing capillary G1362-87300
see"Detector Accessory Kit Contents"
1/4" wrench.
2 Locate the in, waste and recycle ports.
on page 35
Recycle Waste
In
1200 Series RID User Manual