Specifications - DeVilbiss 9000 series Instruction Manual

Nasal cpap system
Table of Contents

Advertisement

Available languages

Available languages

16

SPECIFICATIONS

Technical Specifications
Size .............................................................................4.0"H x 7.5"W x 10.8"D
Weight....................................................................................3.6 lbs. (1.7 kg)
Electrical Requirements (9000/9001) ...................................100-240V~, 50/60 Hz
9001 Only ........................................................................12 VDC @ 3 A max
Maximum Power Consumption ...............................................................65 watts
CPAP Pressure Range......................................................................3-20 cm H
Pressure Delay Time..................................................0, 10, 20, 30 or 45 minutes
Delay Pressure Range........................................3 cm H
Operating Temperature Range ...................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range ....................................5% to 95% RH non-condensing
Operating Atmospheric Conditions .....................................Sea level to 9,000 feet
Storage & Transportation Temperature Range. .........-4°F to 140°F (-20°C to +60°C)
Storage & Transportation Humidity Range................5% to 95% RH non-condensing
Maximum Limited Pressure .................................................................28 cmH
Sound Level ........................................30 dBA (@ 10 cmH
Filter specs ..................................................Standard Filter > 3 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Max flow rates
Model 9000
83 L/m
134 L/m
163 L/m
Pressure Stability
Between maintenance intervals
(maintenance interval = 1 year) Model 9001 = ± 1 cmH
Model 9000 Pressure swings (cmH
2
Breaths per minute
10
@ 7 cmH
O
1.0
2
@ 13 cmH
O
1.0
2
@ 20 cmH
O
1.3
2
A-900
S p e c i f i c a t i o n s
(10.2 cm x 19.6 cm x 26.2 cm)
O to prescription pressure
2
O, 1 meter from CPAP)
2
Model 9001
73 L/m
@ 7 cmH
O (1/3 max pressure)
2
118 L/m
@ 13 cmH
O (2/3 max pressure)
2
164 L/m
@ 20 cmH
O (max pressure)
2
Model 9000 = ± 2 cmH
O
2
O
2
O peak to peak at 500mL tidal volume)
15
20
1.0
1.0
1.0
1.0
1.3
1.4
S p e c i f i c a t i o n s
Model 9001 Pressure swings (cmH
Breaths per minute
@ 7 cmH
O
2
@ 13 cmH
2
@ 20 cmH
2
Warranty.................................................................................Two-year limited
Equipment classification with respect to protection from electric shock........Class II
Degree of protection from electric shock................................Type BF Applied Part
Degree of protection against ingress of liquids.............................................None
O
Equipment not suitable for use in the presence of a flammable anaesthetic mixture
2
with air or with oxygen or nitrous oxide.
Mode of operation ...........................................................................Continuous
Definitions
Display – purpose, type, information displayed. The LCD display shows the following
information in each model:
Model 9000
Set Pres
Altitude
Fine Adj
O
2
Dly Time
Dly Pres
Compl Hrs
Unit Hrs
Firmware
Serial #
Language
Off
Error
O peak to peak at 500mL tidal volume)
2
10
15
1.0
1.0
O
1.0
1.0
O
1.1
1.1
Model 9001 Description
Set Pres
CPAP pressure prescribed for the patient in cmH
Altitude setting one of four ranges
Fine adjustment for fine tuning pressure using an
external manometer
Dly Time
Delay time prescribed for the patient in minutes
Dly Pres
Delay pressure prescribed for the patient in cmH
Compl Hrs
Compliance meter, shows accumulated time "while
breathing"
Unit Hrs
Unit hour meter, shows accumulated time of blower
operation
Firmware
Firmware version installed in the CPAP
Serial #
Serial number of the CPAP
Language
Language selection for LCD display messages
Off
Blower is OFF AC or DC power is connected
Error
An error has been detected by the firmware, the
CPAP is in a fail safe mode
17
20
1.1
1.1
1.2
O
2
O
2
A-900

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

9001 series

Table of Contents