Specifications
from page 13
Continued
14
Measurement time
Agilent 4284A setting time
Integrated time
Short
Medium
Long
Agilent 4285A setting time
Integrated time
Short
Medium
Long
Option 4284A/4285A-001:
Measurement accuracy
The following conditions must be met:
1. Warm up time: ≥ 30 minutes
2. Ambient temperature: 23 °C ±5 °C
3. Test signal voltage: 0.3 Vrms to 1 Vrms
4. Test cable length: 0 m
5. Open and short corrections have been performed
6. D ≤ 0.1 for C, L, X and B measurements
Q ≤ 0.1 for R and G measurements
See operation manual for additional conditions.
Accuracies are relative to calibrated standards. Absolute accuracies are given as:
(Agilent 4284A's relative accuracy +calibration uncertainty of standards).
Accuracy equations
|Z|, |Y|, L, C, R, X, G and B accuracies are given as:
±[A + (K
where:
1. A is basic accuracy as shown in figure 1
2. K
The K
The K
3. K
D accuracy is given as:
±[A
where:
1. A
Q accuracy is given as (when Q
±[(Q
where:
1. Q
2. D
Time interval from a trigger command to the EOM
(end of measurement) signal output at the handler
interface port.
100 Hz
1 kHz
270 ms
40 ms
400 ms
190 ms
1040 ms
830 ms
75 kHz ~ 30 MHz
30 ms
65 ms
200 ms
DC bias current output: 100 mA max
(Agilent 4284A only)
+ K
+ K
) x 100] (% of reading)
a
b
c
and K
are impedance proportional factors given in Table 2.
a
b
term is negligible for impedance above 500 Ω.
a
term is negligible for impedances below 500 Ω.
b
is the calibration interpolation factor given in Table 1.
c
/100] (absolute D value)
e
= [A + (K
+ K
+ K
) x 100]
e
a
b
c
x D
< 1):
x
e
2
x D
)/(1 ](Q
x D
)] (absolute Q value)
x
e
x
e
is the measured Q value
x
is the D accuracy
e
10 kHz
1 MHz
30 ms
30 ms
180 ms
180 ms
820 ms
820 ms
Specifications continued on page 15