Supplemental Performance Characteristics - Agilent Technologies 4284A Technical Overview

Precision lcr meter family 20 hz to 1 mhz 75 khz to 30 mhz
Hide thumbs Also See for 4284A:
Table of Contents

Advertisement

Specifications
from page 12
Continued
General
Power requirements
Power consumption
Operating temperature
and humidity
Size
Weight

Supplemental performance characteristics

Agilent 4284A
Stability
Temperature coefficient
Agilent 4285A
Stability
|X|, |Y|, L, C, R
D
Temperature coefficient
|X|, |Y|, L, C, R
D
Settling time
Frequency
4284A
4285A
Test signal
Range
Input protection
100 V/120 V/220 V ±10%, 240 V +5%/-10%,
47 Hz to 66 Hz.
200 VA
0 °C to 55 °C, < 95% RH at 40 °C
426(W) x 177(H) x 498(D) mm
Agilent 4284A: 15kg (33lb.) Agilent 4285A: 16kg (35.2lb.)
(Not guaranteed)
Medium integration and constant operating temperature
of 23 °C ±5 °C.
|Z|, |Y|, L, C, R < 0.01%/day D < 0.0001/day
Medium integration and 23 °C ±5 °C.
Long integration and constant operating temperature
of 23 °C ±5 °C.
≤ 1 MHz
30 MHz
0.01 %/day
< 0.05 %/day
0.0001/day
< 0.0005/day
Long integration, test signal voltage ≥ 20 mVrms
and 23 °C ±5 °C.
≤ 1 MHz
30 MHz
0.004%/°C
<0.05%/°C
0.00004/°C
<0.0005/°C
≥ 1 kHz
< 70 ms; f
m
< 120 ms; 100 Hz ≤ f
< 1 kHz
m
< 160 ms; f
< 100 Hz
m
< 50 ms
< 120 ms (4284A) < 100 ms (4285A)
≥ 1 kHz
< 50 ms/range shift; f
m
Internal circuit protection, when a charged capacitor is
connected to the unknown terminals. The maximum
capacitor voltage is:
Vmax =√ 1/C (v) where:
C is in Farads
Specifications continued on page 14
Vmax ≤ 200 V
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4285a

Table of Contents