Set The Power Switch To On; Open The Cover - Sanyo MLS-3751 Instruction Manual

High-pressure steam sterilizer
Hide thumbs Also See for MLS-3751:
Table of Contents

Advertisement

2. Set the power switch to ON

• The digital display of the control panel lights up. (The boiling point
setting temperature appears on the digital display I and
on the digital display II for the first 2 seconds.)
Digital display
Cover being CLOSE
Settings (set temperature and sterilization
time/cycle) are indicated.
Digital display I: Sterilization temperature
Digital display II: Sterilization time and cycle
indicated alternately
This status is referred to as
Status .

3. Open the cover.

1) Be sure to check the following points before
opening the cover.
If you attempt to open the cover by force, trouble may
be caused with the product.
• Power switch is turned ON.
• Pressure gauge indicates 0MPa.
• COVER LOCK lamp goes out.
2) While pressing down the handle,
3) Pull the movable handle toward you.
The cover can not be unlocked unless the handle
is pressed down.
4) Open the cover.
To check the temperature inside
the chamber,Press the
and
The
current
indicated while these buttons are
Standby
kept pressed.
Digital display I: Temperature inside
the chamber
Digital display II: Temperature of
heating water
If the temperature is lower than 25°C,
appears.
(The cover can not be opened unless the power switch is turned ON)
- - -
appears
buttons simultaneously.
temperature
is
The figure shown above is for the
case when the boiling point set
temperature is 100°C.
Cover being OPEN
The current temperature of the
heating water in the chamber is
indicated.
The figure shown above is for
the case when the temperature
of the heating water in the
chamber is 30°C.
Make sure that the
COVER LOCK lamp goes out.
Make sure that the pressure
gauge indicates O .
Make sure that the power
switch is turned ON.
Handle
Movable
2)
handle
4)
3)
16