On Power Failure; (For Program Operation) - Sanyo MLS-3751 Instruction Manual

High-pressure steam sterilizer
Hide thumbs Also See for MLS-3751:
Table of Contents

Advertisement

3) Similarly, set the hour and minute in
turn.
4) Press the
button. (Timer operation starts.)
The sterilization temperature appears on
the digital display I and the date and
time appear alternately on the digital
display II.
To cancel the timer operation, press the
button.
Timer setting
Year, month, day, hour and minute have been set at the time of
shipment.
If the timer setting is needed anew after replacement of parts, etc.,
set it in the procedure as follows:
When the pushbuttons
year entry appears and flickers.Enter Christian year by
buttons and press the
Enter the month, day, hour and minute in the same manner.
When setting the minute, press the
by time signal.

On Power Failure

• In power failure, all of the control panel displays are cleared.
• When power failure occurs or when the power switch has been set
to OFF, the safety device works to lock the cover, making it
impossible to open the cover.
To use the product after restoration of power
• When the power is restored, the control panel displays are in standby status.
1) Check the following points before opening the cover.
• Pressure is lowered to 0MPa.
• COVER LOCK lamp goes out.
When the COVER LOCK lamp is lit, the temperature or pressure
inside the chamber rises to make it impossible to pull the movable
handle toward you.
2) Take out the sterilization objects
If liquid boils over in liquid sterilization, clean the inside of the
chamber.
Inside of the chamber : See page 25.
When taking out the sterilization objects, wear heat-resistant
leather gloves for prevention of scald.
3) Check the sterilization by sterilization indicator.
Objects can not be sterilized in some cases.
4) If objects can not be sterilized, repeat the operation.
and then
are pressed, the Christian
button.
button to set exact time
CAUTION
and
Lamp goes out.
24