How to Use

4. Pour the heating water.

• Make sure that the drain valve is closed.
• Set the heater cover in the chamber.
• Pour the water until the tip of the water level bracket of the heater
cover is immersed in the water. Since the amount of water decreases
with every sterilizing operation, check the water level ahead of
operation.
MEMO
The amount of heating water is 3.5 liters.
Advisory note
Purified water is best suited as heating water. If it is unavailable,
use city water instead.
5. Place the sterilization objects inside.
• When using the attached stainless steel baskets in pile, be sure to
place the small basket on the top.
• Set the baskets gently along the guides in the chamber.
• Bring down the grip.
Do not allow the sterilization objects to block the holes or
temperature sensor in the chamber and do not apply undue force
to the temperature sensor.
MEMO
Before, starting operation, check again how to place the sterilization
objects in reference to the On Sterilization Objects (page 12 ~ 13).

6. Close the cover.

1) Make sure on the digital display I that the temperature inside
the chamber is under 60°C
(
appears when the temperature is under 25°C)
MEMO
If the cover is closed when the temperature inside the chamber is over
60°C due to continuous operation, etc., the air inside the chamber will
expand suddenly to make it difficult to open and close the cover. In
this case, wait until the temperature will lower below 60°C.
2) Check
to see whether or not the cover packings and the
chamber opening are free from dirt or dust
If they are dirty or dusty, steam may leak to cause trouble. Keep
them always clean.
(Cover packings and chamber opening
3) Close the cover while bringing down the handle.
Then, a peep sound is heard, and the sterilization set temperature
appears on the digital display I and the sterilization time or the
sterilization time and cycle appear alternately on the digital display II.
17
WARNING
See page 26.)
Drain valve
Drain outlet
Heater cover
Heater cover
Tip (water level)
Heater
Small basket
(Basket without bracket)
Bracket
Small
Large
Handle
Water level
Heating water
Small basket
Small
Large
Large
Grip
Temperature
sensor
Hole
Peep