Before Using The Product - Sanyo MLS-3751 Instruction Manual

High-pressure steam sterilizer
Hide thumbs Also See for MLS-3751:
Table of Contents

Advertisement

Before Using the Product

Installation
• Product environment and conditions of use
Location : Indoor
Ambient temperature : 5 ~ 40°C
Relative humidity : 50 ~ 80%
• The boiling point must be set for the product according to the altitude
of the location where the product is installed. The boiling point has been
set at 100°C at the time of shipment from the factory. Before installing
the product in a place 500m higher than sea level, consult your dealer.
• Do not install the product in the following places. Otherwise, trouble
may occur with your product.
⋅ Places exposed to direct sunlight
⋅ Places where water splashes
⋅ Inclined surfaces (It becomes difficult to detect heating without water.)
⋅ Places exposed to air containing salt, sulfur, etc.
⋅ Humid places
⋅ Dusty places
• Since steam is emitted from the exhaust tank on the rear side of the product, do
not install the product in places where an electric outlet or equipment exists.
• Leave a clearance of more than 15cm behind the product and more
than 15cm on the sides of the product. If the product is installed in
close proximity to walls, the product may overheat, resulting in trouble.
• Wall surfaces may be marred by venting steam or crack under high humidity.
• Anchor the product firmly by means of tip-resistant metals (2
pcs. included with the product).
Three tip-resistant metals are included with the product. Use two of
them for anchoring the product. If the product is not anchored directly
to the floor, anchor it to a floor plate using all three tip-resistant
metals.
• Be sure to install the product on a level rigid surface and lock all four casters.
• Be sure to mount the exhaust tank on the rear side of the product
and the tank mounting brackets to the rear side of the product.
(See page 9.)
Power supply and ground connections
• The unit's power cord must be connected to its own dedicated and
properly wired connection outlet where no other plugs are connected.
For 120V units :
MLS-3750 ;
The unit should be connected to an appropriate receptacle that is rated 16.7 or
greater.
For 220V units :
The 220V versions of the MLS-3751 / 3781 Autoclave are shipped without a
plug on the power cord. Use the power-plug when the products is set. The
protective grounding conductor of power supply cord must be connected to the
contacting member of an attachment plug.
rating ;
Wire size
Power cord
Type
Plug
For 230V units :
The 230V versions of the MLS-3751 / 3781 Autoclave are shipped without a
plug on the power cord. Use the power-plug which is applied IEC60309 when
the products is set. The protective grounding conductor of power supply cord
must be connected to the contacting member of an attachment plug.
rating ;
Power cord Wire size
Plug
• The unit must be connected straight to ground.
The green / yellow conductor of the power cord is the grounding wire. You must
ask your dealer or a qualified electrician to do the grounding work involved.
Altitude
: 0 ~ 3000m
Voltage fluctuation : ± 10% of nominal voltage
M L S - 3 7 5 1
AW G 1 4
S J T
250VAC, 10A or greater
M L S - 3 7 5 1
AW G 1 4
250VAC, 10A or greater
ON
Tip-resistant metal
M L S - 3 7 8 1
250VAC, 20A or greater
M L S - 3 7 8 1
250VAC, 20A or greater
More than 15cm
Screws
(Remove the screws on
the right and left sides
the product, and
retighten them with the
tip-resistant metals.)
Commercially available
anchor bolts (M4 L=20mm
or longer.
Nut-tightening type)
8