Sanyo MLS-3751 Instruction Manual page 4

High-pressure steam sterilizer
Hide thumbs Also See for MLS-3751:
Table of Contents

Advertisement

Safety Precautions
Do not sterilize objects in a steam
impermeable container or bag.
Failure to observe this warning may result in accident
or incomplete sterilization.
Do not block the holes or temperature
sensor in the chamber with objects.
Failure to observe this warning may destabilize control
of the product or result in incomplete sterilization.
When sterilizing only containers such
as beakers, flasks and test tubes,
place them with their openings facing
downward or lay them on their sides.
If the containers are placed with their openings facing
upward, it is hard for air to sufficiently escape from the
container and for steam to sufficiently penetrate inside,
thus resulting in incomplete sterilization.
3
WARNING
Temperature sensor
Hole
Do not seal the object container tightly.
Use a porous cap or loosen the cover sufficiently.
Failure to observe this warning may result in accident
or incomplete sterilization.
Do not place objects on the
temperature sensor in the chamber.
Failure to observe this warning may destabilize control
of the product or result in incomplete sterilization.
When sterilizing waste, pour about
300ml of water into the sterilization
bag and sterilize the waste with the
bag open.
Sterilizing without water in the bag or with the bag
closed may result in incomplete sterilization.
When using the sterilization bag, be
sure to put it in the stainless steel
basket included with the product and
set it in the chamber for sterilization.
If the sterilization bag is put directly in the chamber, it
may block the temperature sensor or it may be pinched
between the chamber opening and the cover packing,
resulting in trouble with the product.
Temperature sensor