3

INSTALLATION

2
1
Fig. 1
Field number
22
Fig. 2
Installing the Camera Head
1
Screw in the U-TV1XC C-mount camera adapter 1 into the mount
thread at the bottom of the camera head 2. If you use a different
C-mount camera adapter, follow its instruction manual.
As the photographed field is as shown below, use a camera
adapter having magnification of 0.63X to 1X. (Vignettings may
occur in 4 corners in the image when the camera adapter
magnification is set to 0.35, or 0.5x depending on the microscope
to be combined.)
If a C-mount TV adapter from other manufacturers than Olympus
is used, the optical performance of the system may not be fully
manifested.
• Be sure to adjust the parfocality before using a camera
0.63x
(FN 17.5)
• Be careful in using other manufacturer's C-mount
1x
(FN 11)
adapter. Otherwise, the focusing of the camera image
will not match that of the image observed through
eyepieces. For the parfocality adjustment method, refer
to the instruction manual for the camera adapter in
use.
camera adapter or C-mount lens
length over 4.5 mm
. Otherwise, the threaded section
b
will hit the inside of the camera head and cause
damage to it.
DP26
a
having a thread
b
a
12