SECTION 3
Operation
WARNING
-/%
NO INTERNAL OPERATOR
Figure 3-2. Model 3321 Rear Panel
CAUTION
FOR FIRE PROTECTION,
SERVICEABLE
PARTS. SERVICE
3.2.2
Rear Panel
CAUTION
0 to *35V is the maximum
voltage that can be
GLIB -
This connector
is used to connect
the GPIB
applied to the BIAS IN connector. Voltages out-
interface
of the instrument
to GPIB interface
of a com-
side this range may blow the fuse and cause
puter. See Section 4 for details.
damage to the instrument.
BIAS -The
following switch, fuse and input connector
pertain to using an external bias supply:
INT/EXT
Switch -
This toggle switch is used to
switch between
the built-in
(INT) bias power sup-
ply (2V) and the externally
(EXT) applied bias sup-
ply (up to 35V).
Fuse -
Maintenance
fuse for the external
power
supply (O.lA quick acting type).
IN Connector
-
BNC connector
used to connect
the
external bias supply to the instrument power input con-
nector. Use a 0 to +35V supply with low ripple
and
noise.
Ventilation Fan -
The fan is used to keep the inside of
the instrument
free from damaging
heat build-up.
The
fan pulls cool air into the instrument
and expels warm
air out of the vents
in the top cover.
For proper
ventilation:
1. Maintain at least 10cm between the rear panel and a
wall.
2. Do not block the air vents in the top cover.
3. Clean the air filter periodically
to prevent clogging.
Figure
3-3 shows
how
to remove
the filter
for
cleaning.
WARNING
Be sure to turn off the power and remove the
power cord before
removing
or installing
the
filter.
3-6