A.5 TRIGGER
Trigger
mode:
Automatic only.
Trigger
delay time: 0
to 199.99s
A.6 MEASUREMENT
TERMINALS
4 terminals (BNC) + guard terminal
A.7 SETUP MEMORY
Memory Content:
All settable data (except bias on-off).
Battery Life: 3
years minimum when stored at 40°C max.
A.8 GPIB
Setting:
Of the items settable via the front panel, all the parameters except
address and delimiter of GI'IB can be set. Also, trigger, OPEN/SHORT
compensation
and memory operation can be performed.
Readout:
All the settable parameters,
measurement
data and status.
Standards:
Based on IEEE-488-1978
and IEEE-488A-1980.
Code:
IS0 7 bit code (ASCII code).
A.9 GENERAL
Power requirements:
AC line voltage: selectable
to lOOV, 12OV, 22OV,
240V ~00% (25OV max.). 48 to 62Hz, approx. 21VA.
Operating Environment:
0" to 4O"C, 10 to 90% RH (non-condensing).
Storage Environment:
-10 to +5O"C, 10 to 80% RH (non-condensing).
Dimensions, Weight:
216mm wide x 132.5mm high
x
350mm deep
(8-l/2 in. x 5-l /4 in. x 13-3/4 in.), excluding protrusions.
Net weight
3.6kg (7.91b.l.
Interface Functions:
SHl, AHl, T6, L4, SRl, RL2, PPO, DCl, DTl, CO.