Braun Perfusor Space Instructions For Use Manual page 81

Hide thumbs Also See for Perfusor Space:
Table of Contents

Advertisement

Chapter 12
Connection Lead SP (12 V) (8713231)
Install the Connection Lead SP (12 V) in the following way:
The green LED of the electronic box shows the operating voltage. The mains
connector can easily be replaced by another plug if required.
Caution:
charging!
Note:
Connection Lead for Staff Call SP (8713232)
To connect the Perfusor® Space to staff call, use the Connection Lead for Staff
Call SP. The staff call needs to comply with the requirements of VDE 0834
(consider country specific regulations).
Note:
The Perfusor® Space offers three different staff call operating modes. They are
displayed in the signalling scheme. Consider the staff call of the hospital when
choosing an operating mode. Choose the operating mode via the service
program.
static
without
Off-
Alarm *
dynamic
without
Off–Alarm
dynamic
with
Off-Alarm
)
*
in the mode static without Off-Alarm, the staff call can be surpressed with
1.) Connect plug to socket P2 on back of pump or F3 on SpaceStation
respectively.
2.) Put the connection lead into the car socket.
3.) If necessary, remove red adaptor of motor vehicle connector by slightly
turning and simultanously pulling.
Do not connect the pump to a patient during external car battery
A maximum of three plugs can be stacked upon each other in socket P2.
Test staff call signalling before every use.
turned off
Alarm
)
Operation
Alarm
Operation
Alarm
Operation
INSTRUCTIONS FOR USE ACCESSORY
turned on
Operating Alarm
1 sec
1 sec
turned off
1 sec
k
81